Operation
Starting Up the System
6-6
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
Starting Up the System
After the system was shut down, the following procedure should be
executed to start the system again.
Initial position:
•
The system is switched off.
•
The system is vented.
❖
To start the instrument
1. Switch the main power switch to position On, see
2. Start the vacuum system in the Instrument Control window
a. Click
in the Vacuum submenu.
b. Wait for the turbomolecular pump to reach at least a frequency
of 800 Hz.
3. Check the vacuum. The vacuum pressure is correct when the
vacuum LED is green, see
.
4. If the system was vented, the detector should be calibrated regardless
of whether it was changed or not.
The Get Ready procedure can now be started, see
.