Operation
System Controls
6-4
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
System Controls
shows the front of an iCAP Q instrument:
System Status LEDs
Three LEDs at the front of the instrument (see
) indicate the
status of the main items of the iCAP Q MS.
meaning.
MS:
Labeled Components: 1=system LEDs
Figure 6-1.
ICP-MS front with system control LEDs
1
Table 6-1.
System Status LEDs of the iCAP Q MS
LED
Status
Information
Power
Green
Main power switch on
Off
Main power switch off
Vacuum
Green
The high vacuum in the analyzer is
reached.
Red
Malfunction of the turbo pump
Orange
The turbo pump is >800 Hz.
Orange flashing
The turbo pump is <800 Hz.
Off
System is vented
System
Blue
System on
Blue flashing
A plasma interlock is active
Off
System off