System Setup
Placing the System
5-4
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
shows the minimum required floor space for the system with
autosampler. All values are given in mm:
Moving the Instrument
The iCAP Q mass spectrometer is provided with four retractable
handles for transport. The instrument is shipped on a pallet with the
handles pulled out, see
. Safety catches prevent the handles
from inadvertently being retracted after they have been pulled out.
Inspect the handles to verify that their safety catches are engaged before
you start lifting the instrument. Once the instrument is in place, push
down the safety catch on a handle to slide it into the instrument.
Labeled Components: 1=monitor, 2=mouse, 3=keyboard, 4=iCAP Q MS, 5=wall, 6=autosampler housing, 7= computer
table
Figure 5-2.
Required floor space for iCAP Q MS with autosampler
700-1000
1400
1391
2100-2400
3500-3800
1000
15
4
5
6
7
1
2
3
773
1015
Figure 5-3.
Retractable handles for transport, fixed on the transport pallet