System Description
Gas Supply
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
3-15
Connection of the laboratory argon supply to the iCAP Q instrument is
made via a push-in fitting. A recommended set-up for the argon supply
is shown in
:
NOTICE
The polyurethane tubing supplied with your iCAP Q system
should not be used between the gas source and the pressure regulator
because polyurethane is not specified for pressure above approximately
0.8 MPa (8 bar).
▲
CCT Gas
Collision Cell Technology (CCT) allows the selective removal of
polyatomic interferences. If an iCAP Q model including CCT is
ordered (iCAP Qc or iCAP Qs), CCT operation at the system is
possible.
NOTICE
The CCT gas outlet should be located within 2 to 3 meters to
the left of the iCAP Q instrument.
▲
The following services must be supplied by the customer:
•
Gas
•
Two-stage pressure regulator
Labeled Components: 1=pressure regulator, 2=shut-off valve, 3=pressure
gauge, 4=argon inlet, 5=iCAP Q, 6=gas source
Figure 3-15.
Argon connection of the iCAP Q instrument
1
2
3
4
5
6