Maintenance
Maintaining the Interface
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
8-19
Cleaning the Sample Cone and the Skimmer Cone
❖
To clean the cone
1. Place the cone in 2% nitric acid in an ultrasonic bath for
15 minutes.
2. Rinse thoroughly with deionized water.
3. Sonicate the cone for 5 minutes with deionized water.
4. Dry it carefully with a non-fiber shedding laboratory wipe or leave it
to air dry.
Inserting the Sample Cone
❖
To insert the sample cone
1. Make sure that the sample cone gasket is in good condition, see
. Replace it if required.
Figure 8-10.
Cone tool, sample cone holder and sample cone
CAUTION
!
Corrosive Acids.
When handling the corrosive acids required to
perform the cleaning procedures described below, take the following
precautions.
•
Wear protective clothing.
•
Use acid resistant laboratory gloves and glasses.
•
Make yourself aware of all safety procedures in case of spillage or
exposure to skin or eyes before commencing the cleaning.
The acid cleaning should take place in a fume hood.