System Setup
Laboratory Conditions
5-8
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
Laboratory Temperature
The mass spectrometer is designed to operate at the temperatures
specified in
NOTICE
As the laboratory temperature increases, system reliability
decreases. All electronic components generate heat while operating. This
heat must be dissipated to the surrounding air for the components to
continue to operate reliably.
▲
Humidity
The relative humidity of the operating environment must be between
20 and 80%, with non-condensing and non-corrosive atmosphere. It is
recommended that your laboratory be equipped with a temperature and
humidity monitor to insure that your laboratory is always within the
required temperature and humidity specifications.
NOTICE
Operating an iCAP Q system at very low humidity might
cause the accumulation and discharge of static electricity, which can
shorten the life of electronic components. Operating the system at high
humidity might cause condensation, oxidation, and short circuits.
▲
Altitude
The iCAP Q instrument is designed for indoor use at an altitude of up
to 2000 m (6562 ft) above sea level. For altitudes above 2000 m, contact
Thermo Fisher Scientific.
Table 5-2.
Temperature requirements for the iCAP Q instrument
Specification
Value
Laboratory temperature
15 to 35 °C