Maintenance
Maintaining the Inlet System
8-14
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
4. Pull off the elbow connector.
5. Remove the drain tubing.
Cleaning the Spray Chamber
For the cleaning procedure of the spray chamber, see
.
Inserting the Spray Chamber
❖
To install the spray chamber
1. Mount the elbow tube to the spray chamber.
2. Attach the drain tubing.
3. Insert the spray chamber with the elbow connector.
4. Mount the Peltier cooling lid.
5. Fit the thumb screw and secure the lid.
6. Insert the nebulizer.
Maintaining the Injector
Removing the Injector
❖
To remove the injector
1. Remove the nebulizer, see
2. Remove the spray chamber, see
3. Carefully remove the injector and the injector holder.
Figure 8-5.
Injector with holder