Maintenance
Servicing Plan
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
8-5
Servicing Plan
Routine and infrequent maintenance procedures to be performed by the
user are listed in
.
Table 8-1.
Maintenance procedures
Item
Procedure
Frequency
Procedure location
Inlet system
•
Peristaltic pump
tubing
Check for clogging or damage
Daily
Replace the tubing
If necessary
•
Nebulizer
Clean the nebulizer
If necessary
Replace the nebulizer
If necessary
•
Spray chamber
Clean the spray chamber
If necessary
Replace the spray chamber
If necessary
•
Injector
Clean the injector
If necessary
Replace the injector
If necessary
•
Torch
Clean the torch
If necessary
Replace the torch
If necessary
Interface
•
Cones
Cleaning the cones
Depending on operation
condition
Replacing the skimmer cone
insert
Depending on operation
condition
•
Extraction lens
Maintaining the lens
Depending on operation
condition
Gas supply
Leak check gas lines
Annually
Air Filter
Check fan
Monthly
Clean filter
If necessary
Instrument
Clean surface
If necessary
Check warning labels
Annually