Maintenance
Maintaining the Interface
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
8-21
d. Carefully pull out the skimmer cone.
4. Remove the skimmer cone from the tool.
5. Remove the insert from the skimmer cone, see
.
Cleaning the Skimmer Cone
For the cleaning procedure of the cone, see
Maintaining the Skimmer Cone Insert
❖
To replace the skimmer cone insert using the insert tool
1. Align the insert tool with the holes in the skimmer cone insert.
Figure 8-12.
Cone tool with skimmer cone
Figure 8-13.
Cone tool and skimmer cone