Operation
Calibration
3-28
410iQ Instruction Manual
Thermo Scientific
The Manual O
2
Calibration screen allows the O
2
span coefficients to be
changed manually while sampling span gas of known concentration. This
button appears if the O
2
Sensor option is selected in the Configuration
screen.
Home Screen>Calibration>Advanced Calibration>Manual Calibration
(single range mode)
The Manual O
2
Calibration screen contains the following information:
●
Adjust O
2
Span Coefficient:
Allows the user to manually adjust the O
2
span coefficient.
●
Reset O
2
Coefficient to 1.000:
Resets coefficient.
Manual O
2
Calibration