Servicing
Optional DMC Oxygen Sensor
7-62
410iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the oxygen sensor board.
Equipment required:
Phillips drive, #2
1.
Remove the oxygen sensor DMC from the instrument. Refer to
“
Oxygen Sensor Removal
” on page
7-59
.
2.
Unplug J1.
3.
Using a #2 Phillips drive, unfasten five #6-32 pan head screws.
Figure 7–71.
Replacing the Oxygen Sensor DMC Board
4.
Slide board upwards.
Note
Clear keyway.
▲
Oxygen Sensor
Board Replacement
Содержание 410iQ