Servicing
Optional DMC Oxygen Sensor
Thermo Scientific
410iQ Instruction Manual
7-59
Use the following procedure to remove the oxygen sensor from the
instrument case. It is easier to do the following by removing the oxygen
sensor first then remove and replace the following as necessary:
●
Oxygen Sensor removal
●
Oxygen Sensor board replacement
●
Oxygen Sensor capillary replacement
●
Oxygen Sensor replacement
Use the following procedure to remove and replace the oxygen sensor.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug DMC cable to oxygen sensor board DMC (J4).
3.
Disconnect plumbing.
4.
Using a #2 Phillips drive, unfasten captive hardware.
Optional DMC
Oxygen Sensor
Oxygen Sensor
Removal
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......