Calibration
Calibration
4-10
410iQ Instruction Manual
Thermo Scientific
Use the following procedure to adjust the span.
1.
Adjust the zero air flow and the CO
2
flow from the standard CO
2
cylinder to provide a diluted CO
2
concentration of approximately 80%
of the upper range limit (URL) of the analyzer.The total air flow must
exceed the total demand of the analyzer connected to the output
manifold to ensure that no ambient air is pulled into the manifold vent.
The exact CO
2
concentration is calculated from:
)
F
+
F
(
)
F
x
]
[CO
(
=
]
[CO
2
CO
D
2
CO
STD
2
OUT
2
Where:
[CO
2
]
OUT
= diluted CO
2
concentration at the output manifold, ppm
[CO
2
]
STD
= concentration of the undiluted CO
2
standard, ppm
F
CO2
= flow rate of CO
2
standard corrected to 25 °C and 760 mmHg, L/min
F
D
= flow rate of dilution air corrected to 25 °C and 760 mmHg, L/min
2.
Allow the instrument to sample this CO
2
concentration standard until
a stable response is obtained.
3.
From the Home Screen, choose Calibration>
Calibrate Span
Coefficient
.
The user sets the span concentration by pressing the Edit Span
Concentration button. The Calculated Span Coefficient button will
show what the span coefficient will be set to if the Calibrate button is
pressed. Pressing the Calibrate button will save the new coefficient and
calibrate the instrument.
4.
Enter the CO
2
calibration gas concentration using the pushbuttons,
and then press
Calibrate
to calibrate the CO
2
reading to the CO
2
calibration gas.
5.
Record the CO
2
concentration and the instrument’s response.
Calibrate Span
Coefficient
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......