Calibration
Calibration
Thermo Scientific
410iQ Instruction Manual
4-9
Use the following procedure to set the CO
2
reading to zero.
1.
Allow sufficient time for the instrument to warm up and stabilize.
2.
Adjust the dilution system of
Figure 4–1
so that zero air alone is
present in the manifold.
Since not all flow controllers have a positive shut off, it might be
necessary to disconnect the CO
2
input line and cap it.
3.
Allow the instrument to sample zero air until a stable reading is
obtained.
4.
From the Home Screen, press Calibration>
Calibrate Background
.
The Target Concentration button will read 0.000. The Calculated
Background button will display the background needed to make the
current CO
2
concentration go to zero.
5.
In the Calibrate Background screen, press
Calibrate
to set the CO
2
reading to zero and to save the new background.
Calibrate
Background
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......