Servicing
DMC Optical Bench
Thermo Scientific
410iQ Instruction Manual
7-33
Use the following procedure to remove the DMC from the instrument
case.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug DMC cable J6 (bottom).
3.
Disconnect plumbing as shown.
4.
Using a #2 Phillips drive, unfasten six captive hardware.
Figure 7–38.
DMC Removal from Instrument (Standard)
Optical Bench
Removal