Calibration
Calibration
4-6
410iQ Instruction Manual
Thermo Scientific
The Multipoint Calibration should be performed initially when the
instrument is received. The user can choose to calibrate with one, two, or
three points. It is suggested to perform a 3-point Multipoint Calibration.
The Multipoint Calibration is divided into three calibration points. The
range is user definable and the following calibration points are
recommended for the user selected range:
●
Point 1: 80% of range
●
Point 2: 50% of range
●
Point 3: 20% or range
Note
If performing a multipoint calibration after a “bad” calibration or
changing ranges, it is recommended to start off with default values for the
Cal point coefficients.
▲
To set default values, from the Home Screen select Calibration>Advanced
Calibration>
Multipoint Calibration
. From Multipoint Calibration, select
Reset Points to Default Values
. The instrument will automatically reset to
default values.
Multipoint
Calibration
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......