Servicing
DMC Optical Bench
Thermo Scientific
410iQ Instruction Manual
7-47
Figure 7–51.
Motor Installation Shaft Alignment
6.
Let instrument sample zero air for about 90 minutes.
7.
From the Home screen, choose Settings>Measurement
Settings>Advanced Measurement Settings>Optical Bench
Settings>
Continue to Initial S/R Calibration
.
8.
Follow steps in Chapter 4, “
Calibration
” for “
Initial S/R
”.