Servicing
Capillary O-Ring Replacement
7-26
410iQ Instruction Manual
Thermo Scientific
7.
Replace capillary and reassemble in reverse order.
Note
Fitting arrangements, number of capillaries and capillary sizes will
vary per instrument configuration.
▲
Use the following procedure to replace the capillary o-rings.
Equipment required:
O-ring pick tool
1.
Using a metal o-ring pick tool, remove the o-ring.
Note
Be careful in not damaging the o-ring walls during this process. Refer
to
Figure 7–29
.
▲
Use the following procedure to replace the power supply.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug all electrical shown J9, J10, J24, J25, J26, and ground.
3.
Unfasten captive hardware.
4.
Slide power supply left, clearing three case floor plate keyways.
Capillary O-Ring
Replacement
Power Supply
Replacement
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......