System Description
Optical Bench DMC
Thermo Scientific
410iQ Instruction Manual
8-3
Figure 8–1.
410iQ Flow Schematic with Filter Wheel Purge (Standard)
Figure 8–2.
410iQ Flow Schematic with Filter Wheel Purge (High Level)
Calibration Note
The stability of the purge gas is very important. Since
this gas is in the same optical path as the sample, changes in the
composition of the purge gas can change the calibration of the instrument.
Therefore, either zero air or a non-toxic dry inert gas, such as nitrogen, is
required as the purge gas.