Servicing
DMC Optical Bench
7-44
410iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the motor.
Equipment required:
Hex drive, 5/32 and 5/64
Phillips drive, #2
1.
Remove optical bench. Refer to “
Optical Bench Removal
” on page
7-
33
.
2.
Remove purge wheel housing. Refer to “
Purge Wheel Housing
Removal
” on page
7-40
.
3.
Remove filter wheel. Refer to “
Filter Wheel Replacement
” on page
7-
42
.
4.
Using 5/64 hex drive, unfasten two #10-32 socket cap head screws.
Figure 7–49.
Motor Removal from Motor Mount pt 1
Motor Replacement