Operation
Settings
Thermo Scientific
410iQ Instruction Manual
3-79
The Range Setting screen defines the concentration range of the analog
outputs. For example, a CO
2
range of 0–500 ppm restricts the analog
output to concentrations between 0 and 500 ppm.
The Range Setting screen shows the current CO
2
range. The range screen is
similar for the single, dual, and auto range modes. The only difference
between the screens are the words “High” or “Low” displayed to indicate
which range is displayed. For more information about the dual and auto
range modes, see “
Range Mode Selection
” on page
3-77
. Pressing Range
Setting, High Range Setting or Low Range Setting, brings up a numeric
keypad whereby the user can select a range.
Settable ranges according to unit selection include:
ppb
200,000
10,000,000 ppb
(Standard)
ppm 200
10,000 ppm (Standard)
% 0.5
25 % (High Level)
μg/m
3
500,000
20,000,000 μg/m
3
(Standard)
mg/m
3
500
20,000 mg/m
3
(Standard)
g/m
3
10
500 g/m
3
(High Level)
Range Setting
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......