Servicing
DMC Optical Bench
7-46
410iQ Instruction Manual
Thermo Scientific
After installing a new motor or filter wheel, use the following procedure to
align the motor and filter wheel.
Equipment required:
Hex drive, 5/32 and 5/64
Thread locker
1.
With the “DIMPLE” of the shaft aligned with the access hole, carefully
slide chopper wheel over motor shaft until set screw lines up with hole.
2.
Apply a light amount of thread locker to #8-32 set screw.
3.
Using 5/64 hex drive tighten set screw slowly while gently rotating the
chopper wheel back and forth. You should feel the set screw engage
with the “DIMPLE” of the shaft. (The set screw for the chopper wheel
is located opposite the curved slot.).
4.
Make sure 10-32 screws are fastened. If motor is replaced, refer to
Figure 7–49
.
5.
After the filter wheel is installed, spin the wheel and observe that it runs
true on the motor shaft.
Motor and Filter
Wheel Alignment
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......