The Range Mode Selection screen is used to switch between the various
range modes: Single, Dual, and Auto Range.
Home Screen>Settings>Measurement Settings>Range Mode Selection
The Range Mode Selection screen contains the following information:
●
Single:
In single range mode, there is one range, one averaging time,
and one span coefficient.
●
Dual:
In the dual range mode, there are two independent analog
outputs. These are labeled simply as the “High Range” and the “Low
Range”. Each channel has its own analog output range setting,
averaging time, and span coefficient.
This enables the sample concentration reading to be sent to the analog
outputs at two different ranges. For example, the low CO
2
analog
output can be set to output concentrations from 0 to 500 ppm and the
high CO
2
analog output set to output concentrations from 0 to 1000
ppm.
In addition, each CO
2
analog output has a span coefficient so that each
range can be calibrated separately. This is especially necessary if the two
ranges are not close to one another. For example, the low CO
2
range is
set to 0–200 ppm and the high CO
2
range is set to 0–10000 ppm.
●
Auto:
The auto range mode switches the CO
2
analog outputs between
high and low range settings, depending on the concentration level. The
high and low ranges are defined in the Range Settings menu.
For example, suppose the low range is set to 500 ppm and the high
range is set to 1000 ppm. Sample concentrations below 500 ppm are
output based on low range selection and sample concentrations above
500 ppm are output based on high range selection. When the low
range is active, the range mode selection status output is at 0 volts.
Range Mode Selection
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......