Servicing
DMC Optical Bench
Thermo Scientific
410iQ Instruction Manual
7-45
5.
Using #2 Phillips drive, unfasten #8-32 pan head screw to gain access
to #8-32 set screw.
6.
Using 5/64 hex drive, loosen #8-32 set screw two or three turns only.
7.
Carefully slide motor out.
8.
Replace motor and align filter wheel and motor. Refer to “
Motor and
Filter Wheel Alignment
” on page
7-46
.
Figure 7–50.
Motor Removal