![Thermo Scientific 410iQ Скачать руководство пользователя страница 121](http://html.mh-extra.com/html/thermo-scientific/410iq/410iq_instruction-manual_1105140121.webp)
Operation
Settings
Thermo Scientific
410iQ Instruction Manual
3-95
The Analog I/O screen allows for configuring the analog inputs/outputs.
This is only visible if selected in Settings>Configuration>
Analog I/O
.
Home Screen>Settings>Communications>Analog I/O
The Analog I/O screen contains the following information:
●
Analog In:
Allows the user to view and calibrate voltage inputs from
external devices.
●
Analog Out (Voltage):
Allows the user to view voltage outputs.
●
Analog Out (Current):
Allows the user to view current (mA) outputs.
●
Analog Out Under/Over Range Enabled/Disabled:
Allows the user to
select whether or not the analog outputs are allowed to exceed the
selected output range.
●
Analog Out Calibration:
Allows the user to calibrate the analog out
voltage and current to account for varying loads attached to the
terminals.
Analog I/O
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......