Calibration
Calibration
Thermo Scientific
410iQ Instruction Manual
4-5
The Initial S/R (Sample/Reference Ratio) is determined at the factory and
should only be adjusted by a qualified technician. The Initial S/R is used to
correct for the slight variations found from one correlation wheel to
another. The only time the Initial S/ R ratio should be changed is when the
correlation wheel is replaced. Zero Air should be sampled when calibrating
the Initial S/R. There are two ways of calibrating the Initial S/R. The user
can manually change the Initial S/R by clicking on the first cell below the
Initial S/R description. The user will be brought to a keypad screen where a
new value can be entered.
However, it is suggested to run an Auto Calibration for the Initial S/R.
This is done by pressing on the cell labeled Start located in the Initial S/R
column. Allow up to 5 minutes for the Initial S/R to be calibrated. The
user can Stop the calibration at any time by pressing the Stop button
located in the Initial S/R column.
The user can set the Initial S/R to factory defaults by pressing on the cell
labeled Default S/R located in the Initial S/R column.
Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue To Initial S/R Calibration
Initial S/R
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......