Operation
Calibration
Thermo Scientific
410iQ Instruction Manual
3-23
The Adjust Span Coefficient screen is used to manually adjust the span
coefficient.
The following screen is shown in single range mode and dual or auto range
mode. In dual or auto range modes, “High Range” or “Low Range”
buttons are displayed to indicate the calibration of the high or low
coefficient. The Adjust High Range Span Coefficient and Adjust Low
Range Span Coefficient screens function the same way.
Home Screen>Calibration>Advanced>Manual Calibration>Adjust
Span Coefficient (single range mode)
Home Screen>Calibration>Advanced>Manual Calibration>Adjust
High Range Span Coefficient (dual or auto range mode)
The Adjust Span Coefficient menu contains the following information:
●
Adjust Span Coefficient:
User manually adjusts span coefficient when in
single range mode.
●
Adjust High Range Span Coefficient:
User manually adjusts high range
span coefficient when in dual or auto range mode.
Adjust Span Coefficient
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......