Calibration
Manual Calibration
Thermo Scientific
410iQ Instruction Manual
4-13
The Manual Calibration menu allows the user to view and manually adjust
the zero background and span coefficient. These are used to correct the
CO
2
readings that the instrument generates using its own internal
calibration data.
Normally, the zero background and span coefficient are calculated
automatically at the Calibrate Background and Calibrate Span Coefficient
described earlier in the chapter. However, the calibration factors can also be
set manually using the functions as described below.
The following screen is shown in single range mode. In dual or auto range
modes, “High Range” or “Low Range” button are displayed to indicate the
calibration of the high or low coefficient. The Adjust High Range Span
Coefficient and Adjust Low Range Span Coefficient screens function the
same way.
Home Screen>Calibration>Advanced Calibration>Manual Calibration
(single range mode)
The CO
2
background is the amount of signal read by the analyzer while
sampling zero air.
The Adjust Background screen is used to perform a manual zero
background calibration of the instrument. As such, the instrument should
sample zero air until stable readings are obtained. The button labeled
Adjust Background allows the user to change zero background. The second
button called Adjusted Concentration shows what the new CO
2
concentration would be based on the changed zero background. Press the
Calibrate button to save the adjusted zero background value.
Manual
Calibration
Adjust Background
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......