![Thermo Scientific 410iQ Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/thermo-scientific/410iq/410iq_instruction-manual_1105140040.webp)
Operation
Calibration
3-14
410iQ Instruction Manual
Thermo Scientific
The Calibrate Span Coefficient menu contains the following information:
●
Edit Span Concentration:
User enters the span concentration when in
single range mode.
●
Edit High Range Span Concentration:
User enters the high range span
concentration when in dual or auto range mode.
●
Edit Low Range Span Concentration:
User enters the low range span
concentration when in dual or auto range mode.
●
Current High Range Concentration:
Read only. Current high range
concentration reading when in dual or auto range mode.
●
Current Low Range Concentration:
Read only. Current low range
concentration reading when in dual or auto range mode.
●
Current Span Coefficient:
Read only. Displays what the current user-set
span coefficient is when in single range mode.
●
Current High Range Span Coefficient:
Displays what the current user-set
high range span coefficient is when in dual or auto range mode.
●
Current Low Range Span Coefficient:
Displays what the current user-set
low range span coefficient is when in dual or auto range mode.
●
Calculated Span Coefficient:
Read only. After the “Edit Span
Concentration” value is entered, the new calculated span coefficient is
displayed when in single range mode.
●
Calculated High Range Span Coefficient:
Read only. After the “Edit
High Range Span Concentration” value is entered, the new calculated
high range span coefficient is displayed when in dual or auto range
mode.
●
Calculated Low Range Span Coefficient:
Read only. After the “Edit Low
Range Span Concentration” value is entered, the new calculated low
range span coefficient is displayed when in dual or auto range mode.
●
Calibrate:
When pressed, updates the coefficient and the concentration
should match the span concentration.
Содержание 410iQ
Страница 1: ...410iQ Instruction Manual Carbon Dioxide Analyzer 117081 00 1Apr2020 ...
Страница 2: ......
Страница 18: ......
Страница 35: ...Operation Instrument Display Thermo Scientific 410iQ Instruction Manual 3 9 ...
Страница 160: ......
Страница 171: ...Calibration Calibration Thermo Scientific 410iQ Instruction Manual 4 11 ...
Страница 180: ...Maintenance Pump Rebuilding 5 4 410iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 188: ......
Страница 197: ...Servicing Fan Replacement Thermo Scientific 410iQ Instruction Manual 7 9 Figure 7 10 Replacing the Fan ...
Страница 231: ...Servicing DMC Optical Bench Thermo Scientific 410iQ Instruction Manual 7 43 Figure 7 48 Filter Wheel ...
Страница 264: ......
Страница 284: ......
Страница 304: ......
Страница 305: ......