Maintenance and Calibration
Auditing the Tilt Sensor
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-17
Figure 5–25.
K0 audit passed prompt
●
The tilt sensor should be audited once a year.
1.
Connect the PDM unit to its charger and start the WinPDM software.
2.
When in the PDM - COM1 screen, select the “Tilt Audit” button to
display the Tilt Audit screen (
Figure 5–26
).
Auditing the Tilt
Sensor
To perform a tilt sensor audit: