Maintenance and Calibration
Performing a Flow Audit
5-10
PDM 3600 Instruction Manual
Thermo Fisher Scientific
Figure 5–14.
Installing the flow/audit/tubing adapter
3.
Connect the PDM unit to the charger and make sure the charger is
connected to your PC.
4.
Open the WinPDM software and connect to the instrument.
5.
Start a primary sample run on the PDM unit (see Chapter 2).
6.
Wait 10 minutes to allow the flow rate to stabilize, then check the flow
rate reading on the flow meter.
7.
Ensure that this value matches the 2.2 lpm flow rate set point for the
PDM unit. If the flow meter reading is within your acceptance criteria
end the sample run, remove the tubing adapter and reinstall the sample
line. If the flow meter is not within your acceptance criteria, a flow
calibration must be performed. See “Performing a Flow Calibration”
later in this chapter.