Maintenance and Calibration
Performing a Flow Calibration
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-39
6.
In the Flow Calibration screen, the Air Heater and TE Heater must
stabilize before a flow calibration can be performed. A
Warning/Confirmation screen displays the “Connect flow meter to cap
lamp tube inlet” message when the flow calibration can begin.
7.
Install the tubing adapter into the sample line (
Figure 5–57
).
8.
Install the other end of the tubing adapter into the tubing that is
attached to the flow meter (
Figure 5–57
).
Figure 5–57.
Install tubing adapter into the sample line
9.
Select the OK button. The Flow Calibration screen will display with
the “Pump Speed” box active (
Figure 5–58
).
Figure 5–58.
Flow calibration screen with pump speed box
Содержание 3600 PDM
Страница 1: ...Model 3600 PDM Personal Dust Monitor Part Number 42 009904 8Oct2013...
Страница 16: ......
Страница 34: ...Basic Operation PDM Display 2 10 PDM 3600 Instruction Manual Thermo Fisher Scientific Figure 2 17 PDM display...
Страница 52: ......
Страница 70: ......