Maintenance and Calibration
Replacing the Cap Lamp
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-31
Figure 5–45.
Mass transducer chamber
Figure 5–46.
Mass transducer O-rings
3.
Using the cotton swab, apply the silicone compound to the O-rings.
4.
Install the PDM cover plate (refer to the instructions earlier in this
chapter for information on opening the case) and perform a leak check.
If the camp lamp assembly malfunctions, the entire cap lamp assembly
should be replaced. Tool/materials needed for this procedure:
Phillips screwdriver
Small flat-head screwdriver
7/64-inch hex wrench
WARNING
Whenever the PDM case is opened for maintenance, you must
perform a case leak check after reassembling the case.
▲
1.
Locate the top cyclone plate on the right-hand side of the PDM unit
and use a 7/64-inch hex wrench to remove the four hex screws from the
top cyclone plate (
Figure 5–47
and
Figure 5–48
).
Replacing the
Cap Lamp
To replace the cap lamp
assembly: