Maintenance and Calibration
Cleaning the Cyclone and Inlet Tubing
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-13
6.
Locate the red cyclone inlet assembly plate gasket (between the top
cyclone plate and the cyclone inlet assembly).
7.
Check the cyclone inlet assembly plate gasket for signs of wear or
damage, and replace the gasket, if necessary (Seal Replacement Kit: 59-
009961).
8.
Using the small flat-head screwdriver, remove the four screws that
secure the top cyclone plate to the inlet tubing (
Figure 5–18
,
Figure 5–
19
and
Figure 5–20
).
9.
Using an alcohol swab, clean the top cyclone plate.
Figure 5–19.
Top cyclone plate with inlet tubing removed
Figure 5–20.
Inlet tubing and gasket