Maintenance and Calibration
Cleaning the Cyclone and Inlet Tubing
5-14
PDM 3600 Instruction Manual
Thermo Fisher Scientific
10.
Check the inlet tubing gasket for signs of wear or damage, and replace
the gasket, if necessary (Seal Replacement Kit: 59-009961). Using
canned air, direct a stream of air up through the inlet tubing (
Figure 5–
21
).
Figure 5–21.
Cleaning inlet tubing
11.
Using an alcohol swab, clean the end of the inlet tubing.
12.
Install the inlet tubing gasket onto the end of the inlet tubing.
13.
Install the top cyclone plate onto the end of the inlet tubing.
14.
Install the top cyclone plate onto the cyclone inlet assembly.
15.
Install the sample line onto the cap lamp.
16.
Install the grit pot.
17.
Install the mass transducer into the PDM unit.
18.
Perform a sample line leak check.