Basic Operation
Connecting to The Charger
Thermo Fisher Scientific
PDM 3600 Instruction Manual
2-5
6.
If the charger and the PDM are both functioning correctly, the LEDs
on the front of the charger will indicate the status of the unit (
Figure 2–
6
).
a) Solid grey (or off): Indicates that the PDM is not properly connected
to the charger, or the charger power is off.
b) Solid green: Indicates that the battery is fully charged.
c) Solid red: Indicates that the battery is being charged.
d) Flashing green: Indicates the battery is at a minimum 80% charged.
e) Flashing red: Indicates a problem with the charger or the connection.
Figure 2–6.
Charger LEDs
7.
Allow the PDM unit to remain connected to a charger for
approximately 6 hours (to fully charge the internal lithium-ion
batteries) before performing your first sample run.
Note
To fully charge the internal lithium-ion batteries, the PDM unit must
be connected to the charger for approximately 6 hours before performing
the first sample run. Charging must be completed in a fresh air
environment.
▲
Note
The batteries will fully charge after approximately 6 hours. However,
as part of an enhanced charging feature, the charging light will flash green
for about two more hours as the charger will continue to provide a trickle
charge to top off the battery.
▲
Содержание 3600 PDM
Страница 1: ...Model 3600 PDM Personal Dust Monitor Part Number 42 009904 8Oct2013...
Страница 16: ......
Страница 34: ...Basic Operation PDM Display 2 10 PDM 3600 Instruction Manual Thermo Fisher Scientific Figure 2 17 PDM display...
Страница 52: ......
Страница 70: ......