PDM 3600 Instruction Manual
5-1
Chapter 5
Maintenance and Calibration
This section describes the procedures involved in verifying the calibration
of the TEOM 3600 PDM Monitor and maintaining the consistent
operation of the hardware. In addition to regularly scheduled maintenance,
the unit may require other maintenance as necessary, including replacing
the batteries, changing the cap lamp or other procedures.
The TEOM 3600 PDM unit requires the following maintenance:
Daily
●
Charge the internal battery pack
. Charge the internal battery
assemblies before each use, or as necessary.
●
Clean the grit pot
. Clean the grit pot after each use, or as necessary.
●
Clean the mass transducer and sample lines
. Replace the TEOM filter
(Section 2) and clean the mass transducer area and sample flow lines
after each use, or as necessary.
Monthly
●
Perform a flow audit
. Perform a flow audit procedure once per month,
or as necessary.
●
Clean the cyclone and inlet tubing
. Clean the cyclone and inlet tubing
once per month, or as necessary.
Annual
●
K0 audit.
Perform an audit of the calibration constant (K0) once a
year, or as necessary.
●
Tilt Audit.
Perform an audit of the tilt sensor once a year, or as
necessary.
CAUTION
TEOM 3600 PDM maintenance should only be performed by
trained, authorized personnel.
▲
Equipment Damage
Do not use acetone to clean any of the parts of the
PDM.
▲