Maintenance and Calibration
Opening the PDM Case
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-21
WARNING
Whenever the PDM case is opened for maintenance, you must
perform a case leak check after reassembling the case.
▲
1.
Remove the mass transducer from the PDM unit and place it on a flat
surface so that the front is accessible.
2.
Using the Phillips screwdriver, remove the screw from location 12 first
(
Figure 5–31
). Next, in order, remove screws from locations 14, 13, 11,
and 10. Finally, remove the remaining screws in any order of preference
except locations 1 and 2 which should be removed last.
Figure 5–31.
Front cover screws
Figure 5–32.
Screws on top display
Equipment Damage
Do not attempt to remove the security (non-Phillips
head) screws on the top or front of the PDM.
▲
3.
Remove the cover panel (
Figure 5–33
).
To open the PDM case: