Maintenance and Calibration
Replacing the Cap Lamp
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-33
4.
Remove the PDM cover plate (refer to the instructions earlier in this
chapter for information on opening the case).
5.
Locate the red power (supply) wire and open the blue connector on the
red power wire by unscrewing it (
Figure 5–50
).
Figure 5–50.
Red power supply wire
6.
Remove the small blue connector that is attached to the red power line
on the cap lamp assembly-side of the wire (
Figure 5–51
).
Note
: There is a small plastic ferrule located inside the small blue
connector. Be sure to retain the plastic ferrule and blue connector to use
during the installation procedure.
▲
Figure 5–51.
Blue connector removed from camp lamp-side of power wire
7.
Lift up the black insulating material from the cyclone area to display
the wiring underneath (
Figure 5–52
).