Introduction
Advanced Features
1-2
PDM 3600 Instruction Manual
Thermo Fisher Scientific
The user interface includes membrane switches and a top-mounted
character display that provides shift exposure information and status codes.
The high contrast display provides basic data and statistics for the system,
including status conditions. An additional PC-based user interface is
provided through Thermo Scientific’s WinPDM communication software.
The TEOM 3600 PDM is powered by two internal lithium-ion battery
assemblies. See
Table 1–1
and
Table 1–2
for the TEOM 3600 PDM
system specifications.
Table 1–1.
TEOM 3600 PDM Specifications
TEOM 3600 PDM Specifications
Flow Rate
2.2 l/min ±2.5% when calibrated at set point
Run Time: PDM
Minimum of 12 hours
Run Time: Cap Lamp
Minimum of 12 hours while powering with auxiliary
connector at 200 mA, 7.2 VDC
Humidity: PDM
0-100% RH, condensing and non-condensing
Humidity: Charging Station
5-95% RH, non-condensing
Altitude
-3,500 to 10,000 feet
Temperature: Sampled Air
-20º C to 40° C, typical range is 0-30° C
Temperature Shock
40° C/min
Temperature: Charging
0º C to 40º C
Table 1–2.
Temperature of Mass Transducers and Heaters
Ambient Operating
Temperature
Air Inlet Heater
Set Point (C) (Az)
TE Heater
Set Point (C) (Tz)
98-104° F (37-40° C)
57° C
60° C
86-97° F (30-36° C)
50° C
53° C
73-85° F (23-29° C)
43° C
46° C
60-72° F (16-22° C)
36° C
39° C
48-59° F (9-15° C)
29° C
32° C
35-47° F (2 - 8° C)
22° C
25° C
23-34° F (-5 - 1° C)
15° C
18° C
10 - 22° F (-12 - -6° C)
8° C
11° C
-2 - 9° F (-19 - -13° C)
1° C
4° C