Maintenance and Calibration
Repairing/Rebuilding the Cap Lamp
5-36
PDM 3600 Instruction Manual
Thermo Fisher Scientific
Figure 5–54.
Installing the ferrule and connector
17.
Install the blue connectors on the red power wires by screwing them
together firmly.
18.
Reinstall the black insulating material onto the cyclone area to cover
the wiring underneath.
19.
Install the PDM cover plate (refer to the instructions earlier in this
chapter for information on opening the case).
20.
Turn on the cap lamp to ensure that it is functioning.
21.
Perform a leak check.
Use the following procedure to repair/rebuild the cap lamp. Tool/materials
needed for this procedure:
Phillips screwdriver
Slotted screwdriver
7/64-inch Allen wrench
●
Cap lamp kit (p/n 113326-00). Includes:
o
Lens, headpiece
o
Gasket, headpiece
Repairing/
Rebuilding
the Cap
Lamp