Maintenance and Calibration
Replacing the Cap Lamp
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-35
9.
Using a small flathead screwdriver, loosen the screw on the terminal
block that is toward the bottom of the unit and remove the black
return wire from the block (
Figure 5–53
).
10.
Remove the entire cap lamp assembly from the PDM unit.
11.
Pull the red power wire and the black return wire on the new cap lamp
assembly through the opening on the PDM unit for the cyclone inlet
assembly.
12.
Reinstall the bottom cyclone plate to the PDM unit using the three
screws that were removed earlier.
13.
Reinstall the top cyclone plate to the bottom cyclone plate using the
four screws that were removed earlier.
14.
Insert the return wire into the bottom port on the terminal block and
tighten the bottom screw on the terminal block to secure the return
wire.
15.
Pull the red power wire out from behind the cyclone inlet area.
16.
Remove the small plastic ferrule from the small blue connector. Install
the small blue connector onto the red power wire first, and then install
the small plastic ferrule onto the power wire (
Figure 5–54
). Be sure to
install the small plastic ferrule close to, but not on top of, the bare end
of the red power wire.
Screw