Maintenance and Calibration
Replacing the Battery Packs
5-28
PDM 3600 Instruction Manual
Thermo Fisher Scientific
The TEOM 3600 PDM has two battery pack assemblies that should be
replaced when their capacity diminishes. Tool/materials needed for this
procedure include a Phillips screwdriver.
WARNING
Whenever the PDM case is opened for maintenance, you must
perform a case leak check after reassembling the case.
▲
CAUTION
Refer to Section 1 for a comprehensive list of cautions and
warnings for the battery pack assemblies.
▲
Note
. This procedure requires removal of the front panel of the unit.
While the case is disassembled, you should check/lubricate the mass
transducer O-rings. Refer to “Lubricating the Mass Transducer O-rings”
later in this section for more information.
1.
Remove the PDM cover plate (refer to the instructions earlier in this
chapter for information on opening the case).
2.
Locate the two battery pack assemblies in the bottom of the unit
(
Figure 5–42
).
Figure 5–42.
PDM with cover panel removed
Replacing the
Battery Packs
To replace the battery pack
assemblies:
A
B