Maintenance and Calibration
Installing the PDM Cover Panel
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-23
Figure 5–34.
Cover gasket checkpoints
4.
Install the screws in locations 1 and 2 until seated (
Figure 5–35
). Press
down on back of cover, aligning holes 3 and 4. Put a screw into either
location 3 or 4 and tighten lightly. Then torque the screws to 6-in. lbs,
respectively on location 1 and 2.
5.
Install the remaining screws into the front panel to get them started in a
cross-pattern as shown in
Figure 5–35
. Then, tighten until seated in a
cross-pattern as shown in
Figure 5–35
. Finally, torque the screws to 6-
in. lbs in a cross-pattern as shown in
Figure 5–35
.
Figure 5–35.
Front cover screws
6.
Perform a case leak check.