Maintenance and Calibration
Performing Sample Path and Case Leak Checks
Thermo Fisher Scientific
PDM 3600 Instruction Manual
5-27
Figure 5–41.
Leak check screen with a “passed” message
8.
If the leak check passes, remove the case leak check suction cup from
the battery compartment vent and remove the case leak check suction
cup from the sample line. Return to normal operation. If the leak check
fails, refer to the Troubleshooting section.
WARNING
Do not operate the TEOM 3600 PDM if its case is damaged
or otherwise compromised.
▲