UM-BIOIIADVANCE-EN-1302
32
No.
Type of
error
Causes of error
Correction of error
3
Excessive
laminar
down flow
Alarm “
HIGH
LAM. FLOW
”
(see Chapter
4.5 point D)
Probe not working properly.
Check the probe position is correct (sensor
parallel to the flow).
Air currents in front
aperture.
Check and remove any airflow in front of
the cabinet that distorts the reading of the
sensor.
4
Laminar
down flow
not working
properly
Alarm “
LOW
LAM. FLOW
”
(see Chapter
4.5 point D)
Probe not working properly.
Check the probe position is correct (sensor
parallel to the flow).
Check and remove any obstruction to the
probe.
Check that the connection in the control
card is correct (
Annex A7
).
Laminar down flow fan
stopped.
Check the fuse has not blown (
Annex
A7
).
Check that the connection in the control
card is correct (
Annex A7
).
Notify the Technical Service to change the
fan.
5
Exhaust not
working
properly
Alarm “
LOW
EXH. FLOW”
(see Chapter
4.5 point D)
Exhaust completely covered
up.
Visually check the exhaust to remove any
type of obstruction.
Probe not working properly
Check the probe position is correct (sensor
parallel to the flow).
Check that the connection in the control
card is correct (
Annex A7
).
Exhaust fan stopped.
Check the fuse has not blown (
Annex
A7
).
Check that the connection in the control
card is correct (
Annex A7
).
Notify the Technical Service to change the
fan.
6
Front does
not move
Front is blocked.
Visually check the front movement to
remove any type of obstruction.
7
Glass out of
position
Alarm
“
WORK
POSITION”
(see Chapter
4.5 pto. D)
Glass position
Assemble the glass in working position.
Working position micro is
loose
Open the frontal and check that the
working position micro is not loose.
Notify the Technical Service to receive
instructions for adjusting the micro.
Main control is damaged
Call a specialised technician to replace the
control.
Содержание Bio II Advance
Страница 2: ...MANUAL DE USUARIO Cabinas de Seguridad Microbiol gica Bio II Advance...
Страница 3: ...UM BIOIIADVANCE SP 1302 2...
Страница 35: ...UM BIOIIADVANCE SP 1302 34 11 CERTIFICADOS 11 1 CE Ver Anexo A10 1 11 2 GARANT A Ver Anexo A10 2...
Страница 36: ...USER S MANUAL Microbiological Safety Cabinet Bio II Advance...
Страница 37: ...UM BIOIIADVANCE EN 1302 2...
Страница 69: ...UM BIOIIADVANCE EN 1302 34 11 CERTIFICATES 11 1 CE See Annex A10 1 11 2 GUARANTEE See Annex A10 2...
Страница 70: ...MANUEL D UTILISATION Cabines de s curit microbiologique Bio II Advance...
Страница 71: ...UM BIOIIADVANCE FR 1302 2...
Страница 103: ...UM BIOIIADVANCE FR 1302 34 11 CERTIFICATS 11 1 CE Voir Annexe A10 1 11 2 GARANTIE Voir Annexe A10 2...
Страница 104: ...BEDIENUNGSANLEITUNG Mikrobiologische Sicherheitswerkb nke Bio II Advance...
Страница 105: ...UM BIOIIADVANCE DE 1302 2...
Страница 120: ...UM BIOIIADVANCE DE 1302 17 Abb 1 X Abb 2 X Abb 3 X Abb 4 X...
Страница 139: ...UM BIOIIADVANCE DE 1302 36 11 ZERTIFIKATE 11 1 CE Siehe Anhang A10 1 11 2 GARANTIE Siehe Anhang A10 2...
Страница 140: ...ANNEXOS ANNEXES ANNEXES ANH NGE Bio II Advance...
Страница 141: ...A1 BIOIIADVANCE SP EN 1302 1 Vlam Vent Qext SP Anexo A1 ESQUEMA DE PRINCIPIO EN Annex A1 PRINCIPLE SCHEME...
Страница 142: ...A1 BIOIIADVANCE FR DE 1302 1 Vlam Vent Qext FR Annexe A1 SCH MA DE PRINCIPE DE Anhang A1 FUNKTIONSPRINZIP...
Страница 156: ...A5 1 BIOIIADVANCE FR DE 1302 3 Fig 3...
Страница 183: ...A7 BIOIIADVANCE SP EN FR DE 1302 2...
Страница 184: ...A7 BIOIIADVANCE SP EN FR DE 1302 3...
Страница 185: ...A7 BIOIIADVANCE SP EN FR DE 1302 4...
Страница 186: ...A7 BIOIIADVANCE SP EN FR DE 1302 5...
Страница 187: ...A7 BIOIIADVANCE SP EN FR DE 1302 6...
Страница 188: ...A7 BIOIIADVANCE SP EN FR DE 1302 7...
Страница 189: ...A7 BIOIIADVANCE SP EN FR DE 1302 8...
Страница 190: ...A7 BIOIIADVANCE SP EN FR DE 1302 9...
Страница 191: ...A7 BIOIIADVANCE SP EN FR DE 1302 10...
Страница 226: ......
Страница 227: ......