UM-BIOIIADVANCE-EN-1302
31
10.
TROUBLESHOOTING
No.
Type of
error
Causes of error
Correction of error
1
Insufficient
exhaust
flow
Alarm “
LOW
EXH. FLOW”
(see Chapter
4.5 point D)
Exhaust partially covered.
Visually check the exhaust to remove any
type of obstruction.
Front perforation in the
working area
totally/partially covered.
Visually check the perforation and remove
any type of obstruction.
Distance between cabinet
exhaust and lab ceiling too
small.
Increase space between cabinet exhaust
and the ceiling. In some cases it is
possible to modify or eliminate the false
ceiling.
Install a lower support table. This means
the working area will be also lower.
Duct the cabinet (
Chapter 3.5
).
Clogged exhaust filter.
Check filter clogging with the cabinet in
normal operation (
Chapter 4.5 point 2
).
Check the date of the last filter change
and the hours it has been under operation.
(
Chapter 4.5 point 3.3.1
).
Notify the technical service to change the
filter.
Probe not working properly. Check the probe position is correct.
Ducted cabinet and ducting
too long.
Fit an additional fan at the end of the duct
(
Chapter 3.5
).
2
Insufficient
laminar
down flow
Alarm “
LOW
LAM. FLOW
”
(see Chapter
4.5 point D)
Front and rear perforation
in the work zone
totally/partially covered.
Visually check the perforation and remove
any type of obstruction.
Clogged laminar down flow
filter.
Check filter clogging with the cabinet in
normal operation (
Chapter 4.5 point 2
).
Check the date of the last filter change
and the hours it has been under operation.
(
Chapter 4.5 point 3.3.1
).
Notify the technical service to change the
filter.
Probe not working properly.
Check the probe position is correct (sensor
parallel to the flow).
Check and remove any obstruction to the
probe.
Содержание Bio II Advance
Страница 2: ...MANUAL DE USUARIO Cabinas de Seguridad Microbiol gica Bio II Advance...
Страница 3: ...UM BIOIIADVANCE SP 1302 2...
Страница 35: ...UM BIOIIADVANCE SP 1302 34 11 CERTIFICADOS 11 1 CE Ver Anexo A10 1 11 2 GARANT A Ver Anexo A10 2...
Страница 36: ...USER S MANUAL Microbiological Safety Cabinet Bio II Advance...
Страница 37: ...UM BIOIIADVANCE EN 1302 2...
Страница 69: ...UM BIOIIADVANCE EN 1302 34 11 CERTIFICATES 11 1 CE See Annex A10 1 11 2 GUARANTEE See Annex A10 2...
Страница 70: ...MANUEL D UTILISATION Cabines de s curit microbiologique Bio II Advance...
Страница 71: ...UM BIOIIADVANCE FR 1302 2...
Страница 103: ...UM BIOIIADVANCE FR 1302 34 11 CERTIFICATS 11 1 CE Voir Annexe A10 1 11 2 GARANTIE Voir Annexe A10 2...
Страница 104: ...BEDIENUNGSANLEITUNG Mikrobiologische Sicherheitswerkb nke Bio II Advance...
Страница 105: ...UM BIOIIADVANCE DE 1302 2...
Страница 120: ...UM BIOIIADVANCE DE 1302 17 Abb 1 X Abb 2 X Abb 3 X Abb 4 X...
Страница 139: ...UM BIOIIADVANCE DE 1302 36 11 ZERTIFIKATE 11 1 CE Siehe Anhang A10 1 11 2 GARANTIE Siehe Anhang A10 2...
Страница 140: ...ANNEXOS ANNEXES ANNEXES ANH NGE Bio II Advance...
Страница 141: ...A1 BIOIIADVANCE SP EN 1302 1 Vlam Vent Qext SP Anexo A1 ESQUEMA DE PRINCIPIO EN Annex A1 PRINCIPLE SCHEME...
Страница 142: ...A1 BIOIIADVANCE FR DE 1302 1 Vlam Vent Qext FR Annexe A1 SCH MA DE PRINCIPE DE Anhang A1 FUNKTIONSPRINZIP...
Страница 156: ...A5 1 BIOIIADVANCE FR DE 1302 3 Fig 3...
Страница 183: ...A7 BIOIIADVANCE SP EN FR DE 1302 2...
Страница 184: ...A7 BIOIIADVANCE SP EN FR DE 1302 3...
Страница 185: ...A7 BIOIIADVANCE SP EN FR DE 1302 4...
Страница 186: ...A7 BIOIIADVANCE SP EN FR DE 1302 5...
Страница 187: ...A7 BIOIIADVANCE SP EN FR DE 1302 6...
Страница 188: ...A7 BIOIIADVANCE SP EN FR DE 1302 7...
Страница 189: ...A7 BIOIIADVANCE SP EN FR DE 1302 8...
Страница 190: ...A7 BIOIIADVANCE SP EN FR DE 1302 9...
Страница 191: ...A7 BIOIIADVANCE SP EN FR DE 1302 10...
Страница 226: ......
Страница 227: ......