UM-BIOIIADVANCE-EN-1302
10
g)
Due to the ceiling temperature affects air distribution, sites should be selected with
enough space. We recommend 250 mm free space between the exhaust and the ceiling.
h)
The location of the cabinet will directly influence the noise level. This level will rise
considerably in small spaces with very reflective surfaces because of the proximity of the
cabinets to the walls, and where fan speed has risen to compensate for air currents in
the room or the resistance of clogged filters.
i)
The location should allow disinfection gases with formaldehyde (
Annex A5.3
) to be
expelled into the atmosphere (in case that it is allowed by the local regulations of the
country or customer facilities).
j)
Correct cabinet performance should be tested before putting it into operation if any
modification has been made to the room or to the air intake system which may
considerably change the airflow characteristics in the room.
3.3.
PRECAUTIONS FOR FIRST CONNECTION
•
Before connecting it to the mains supply, proceed to make a general cleaning to eliminate
any dust particles that may have accumulated during transportation. Follow the
instructions on cleaning and disinfection.
•
Before using it for the first time, run the cabinet for around 6 hours to purge the filters.
New filters give off a characteristic smell which tends to disappear in a short period of
time.
WARNING
The cabinet should work at the voltage shown on the name plate. It should be connected to
a suitable socket base (minimum 16A) with an earthed plug. The cable section recommended
for the base is 1.5 mm.
WARNING
The plug connection must be located in such a place to prevent it from being unplugged
accidentally during operation, e.g., next to a wall or closed to another equipment.
EARTH
WIRE
Содержание Bio II Advance
Страница 2: ...MANUAL DE USUARIO Cabinas de Seguridad Microbiol gica Bio II Advance...
Страница 3: ...UM BIOIIADVANCE SP 1302 2...
Страница 35: ...UM BIOIIADVANCE SP 1302 34 11 CERTIFICADOS 11 1 CE Ver Anexo A10 1 11 2 GARANT A Ver Anexo A10 2...
Страница 36: ...USER S MANUAL Microbiological Safety Cabinet Bio II Advance...
Страница 37: ...UM BIOIIADVANCE EN 1302 2...
Страница 69: ...UM BIOIIADVANCE EN 1302 34 11 CERTIFICATES 11 1 CE See Annex A10 1 11 2 GUARANTEE See Annex A10 2...
Страница 70: ...MANUEL D UTILISATION Cabines de s curit microbiologique Bio II Advance...
Страница 71: ...UM BIOIIADVANCE FR 1302 2...
Страница 103: ...UM BIOIIADVANCE FR 1302 34 11 CERTIFICATS 11 1 CE Voir Annexe A10 1 11 2 GARANTIE Voir Annexe A10 2...
Страница 104: ...BEDIENUNGSANLEITUNG Mikrobiologische Sicherheitswerkb nke Bio II Advance...
Страница 105: ...UM BIOIIADVANCE DE 1302 2...
Страница 120: ...UM BIOIIADVANCE DE 1302 17 Abb 1 X Abb 2 X Abb 3 X Abb 4 X...
Страница 139: ...UM BIOIIADVANCE DE 1302 36 11 ZERTIFIKATE 11 1 CE Siehe Anhang A10 1 11 2 GARANTIE Siehe Anhang A10 2...
Страница 140: ...ANNEXOS ANNEXES ANNEXES ANH NGE Bio II Advance...
Страница 141: ...A1 BIOIIADVANCE SP EN 1302 1 Vlam Vent Qext SP Anexo A1 ESQUEMA DE PRINCIPIO EN Annex A1 PRINCIPLE SCHEME...
Страница 142: ...A1 BIOIIADVANCE FR DE 1302 1 Vlam Vent Qext FR Annexe A1 SCH MA DE PRINCIPE DE Anhang A1 FUNKTIONSPRINZIP...
Страница 156: ...A5 1 BIOIIADVANCE FR DE 1302 3 Fig 3...
Страница 183: ...A7 BIOIIADVANCE SP EN FR DE 1302 2...
Страница 184: ...A7 BIOIIADVANCE SP EN FR DE 1302 3...
Страница 185: ...A7 BIOIIADVANCE SP EN FR DE 1302 4...
Страница 186: ...A7 BIOIIADVANCE SP EN FR DE 1302 5...
Страница 187: ...A7 BIOIIADVANCE SP EN FR DE 1302 6...
Страница 188: ...A7 BIOIIADVANCE SP EN FR DE 1302 7...
Страница 189: ...A7 BIOIIADVANCE SP EN FR DE 1302 8...
Страница 190: ...A7 BIOIIADVANCE SP EN FR DE 1302 9...
Страница 191: ...A7 BIOIIADVANCE SP EN FR DE 1302 10...
Страница 226: ......
Страница 227: ......