GENERAL WORKING RECOMMENDATIONS
a)
We recommend starting up the cabinet around 10 minutes before you begin to work.
This will ensure particles are swept from the work zone and from material introduced.
b)
Wash your arms, hands and nails well, using an anti-bacterial soap, before and after
work. Staff should avoid touching their mouths and eyes.
c)
We recommend wearing lab coats with adjusted cuffs and, in special works, protection
gloves (or sleeve cuffs). Both the coat and cuffs should be made from a fabric that
minimises the emission of fibres and particles.
d)
The work area inside the cabinet should not be used to store lab equipment. This can
produce an accumulation of dust that is unnecessary and hazardous for working under
sterile conditions. Only the fewest pieces of equipment should be placed in order to work
correctly and with the minimum number of interruptions for the user (inserting and
removing arms from the work zone).
e)
Although the work zone is under a clean and sterile laminar down flow, the area around
the cabinet might be contaminated. You should know the sources of contamination
should be known in order to avoid them when inserting arms, material, etc.
f)
All the material needed for the work should be free of particles and carefully cleaned
before it is placed on the work area.
g)
Materials such as paper, wood, cardboard, pencils, rubbers, etc. should not be placed in
the work area as they shed a lot of particles.
h)
In the event of using pipettes, only mechanical aspiration type should be used. Blowing
the pipette with the mouth should never be done because it is easy to inhale aerosols
caused by suction.
i)
When platinum handles are used, it is advisable to use electric incinerators, preferably
single-use ones.
j)
If the work requires the use of a gas flame from a Bunsen burner or similar, we
recommend using a type where the flame is maintained by holding down a button. Please
note that the use of a constant flame burner creates significant turbulence. Bear in mind
that a flame too large could burn the absolute filters.
k)
Shadows and turbulence provoked by objects, equipment and materials placed in the
work area should be studied before beginning to work in order to evaluate their possible
effects. Bear in mind that the laminarity of the airflow is not recovered until a distance of
2.5 times the diameter of the object that causes the obstruction.
l)
When using bottles and tubes, they should preferably have a screw top instead of cotton
stoppers which easily shed a lot of particles.
m)
When the work is finished, all disposable products (handles, Petri dishes, etc.) as well as
means of cultivation, samples, tubes, bottles, etc. should be removed from the cabinet in
impermeable bags suitable for sterilisation if necessary.
n)
When working in the cabinet and during cleaning operations, it is important to protect
the HEPA filters from any damage by avoiding knocking them, projecting liquids or
splashing, etc.
o)
When placing or taking out the arm rest, take care not to hit strongly the lateral glasses,
in order to prevent any damage of them.
Содержание Bio II Advance
Страница 2: ...MANUAL DE USUARIO Cabinas de Seguridad Microbiol gica Bio II Advance...
Страница 3: ...UM BIOIIADVANCE SP 1302 2...
Страница 35: ...UM BIOIIADVANCE SP 1302 34 11 CERTIFICADOS 11 1 CE Ver Anexo A10 1 11 2 GARANT A Ver Anexo A10 2...
Страница 36: ...USER S MANUAL Microbiological Safety Cabinet Bio II Advance...
Страница 37: ...UM BIOIIADVANCE EN 1302 2...
Страница 69: ...UM BIOIIADVANCE EN 1302 34 11 CERTIFICATES 11 1 CE See Annex A10 1 11 2 GUARANTEE See Annex A10 2...
Страница 70: ...MANUEL D UTILISATION Cabines de s curit microbiologique Bio II Advance...
Страница 71: ...UM BIOIIADVANCE FR 1302 2...
Страница 103: ...UM BIOIIADVANCE FR 1302 34 11 CERTIFICATS 11 1 CE Voir Annexe A10 1 11 2 GARANTIE Voir Annexe A10 2...
Страница 104: ...BEDIENUNGSANLEITUNG Mikrobiologische Sicherheitswerkb nke Bio II Advance...
Страница 105: ...UM BIOIIADVANCE DE 1302 2...
Страница 120: ...UM BIOIIADVANCE DE 1302 17 Abb 1 X Abb 2 X Abb 3 X Abb 4 X...
Страница 139: ...UM BIOIIADVANCE DE 1302 36 11 ZERTIFIKATE 11 1 CE Siehe Anhang A10 1 11 2 GARANTIE Siehe Anhang A10 2...
Страница 140: ...ANNEXOS ANNEXES ANNEXES ANH NGE Bio II Advance...
Страница 141: ...A1 BIOIIADVANCE SP EN 1302 1 Vlam Vent Qext SP Anexo A1 ESQUEMA DE PRINCIPIO EN Annex A1 PRINCIPLE SCHEME...
Страница 142: ...A1 BIOIIADVANCE FR DE 1302 1 Vlam Vent Qext FR Annexe A1 SCH MA DE PRINCIPE DE Anhang A1 FUNKTIONSPRINZIP...
Страница 156: ...A5 1 BIOIIADVANCE FR DE 1302 3 Fig 3...
Страница 183: ...A7 BIOIIADVANCE SP EN FR DE 1302 2...
Страница 184: ...A7 BIOIIADVANCE SP EN FR DE 1302 3...
Страница 185: ...A7 BIOIIADVANCE SP EN FR DE 1302 4...
Страница 186: ...A7 BIOIIADVANCE SP EN FR DE 1302 5...
Страница 187: ...A7 BIOIIADVANCE SP EN FR DE 1302 6...
Страница 188: ...A7 BIOIIADVANCE SP EN FR DE 1302 7...
Страница 189: ...A7 BIOIIADVANCE SP EN FR DE 1302 8...
Страница 190: ...A7 BIOIIADVANCE SP EN FR DE 1302 9...
Страница 191: ...A7 BIOIIADVANCE SP EN FR DE 1302 10...
Страница 226: ......
Страница 227: ......