OPERATING INSTRUCTIONS
4.1.
START-UP
Once the above points have been controlled, start up the MSC by pressing the On/Off button
(
Annex A3, pos. 1
). After entering the password correctly (by default the password is
0000
) or immediately if this function has been disabled, the fans will automatically connect.
Ensure that the aspiration area and air exhaust outlet are free (
Annex A2, pos. 8
).
The electronic system of regulation by microprocessor via a thermoanemometer probe,
laminar down flow and exhaust airflow, needs a few seconds to stabilise during connection in
order to begin to work normally.
The cabinet can only be used by qualified staff who meets the correct work techniques in the
laboratory. If these techniques are not fulfilled, faulty working of the equipment may
happen.
WARNING
The GS certification does not allow the deactivation of the password enquiry.
4.2.
FRONT PANEL POSITIONS
By means of the handles “A”, the front glass can be moved from working position to closed
position or vice versa. To move the glass just slide down or up the glass.
FROM WORKING POSITION
FROM CLOSED POSITION
In case it is needed to clean the glass by the inner part (Annex A5.2) or if it is necessary to
place in the working area any big product or device, the glass-frontal group may be
separated from the cabinet, providing full access to both the inner glass area and the
working area. All needed precautions should be taken as the user protection decreases
during that operation.
Содержание Bio II Advance
Страница 2: ...MANUAL DE USUARIO Cabinas de Seguridad Microbiol gica Bio II Advance...
Страница 3: ...UM BIOIIADVANCE SP 1302 2...
Страница 35: ...UM BIOIIADVANCE SP 1302 34 11 CERTIFICADOS 11 1 CE Ver Anexo A10 1 11 2 GARANT A Ver Anexo A10 2...
Страница 36: ...USER S MANUAL Microbiological Safety Cabinet Bio II Advance...
Страница 37: ...UM BIOIIADVANCE EN 1302 2...
Страница 69: ...UM BIOIIADVANCE EN 1302 34 11 CERTIFICATES 11 1 CE See Annex A10 1 11 2 GUARANTEE See Annex A10 2...
Страница 70: ...MANUEL D UTILISATION Cabines de s curit microbiologique Bio II Advance...
Страница 71: ...UM BIOIIADVANCE FR 1302 2...
Страница 103: ...UM BIOIIADVANCE FR 1302 34 11 CERTIFICATS 11 1 CE Voir Annexe A10 1 11 2 GARANTIE Voir Annexe A10 2...
Страница 104: ...BEDIENUNGSANLEITUNG Mikrobiologische Sicherheitswerkb nke Bio II Advance...
Страница 105: ...UM BIOIIADVANCE DE 1302 2...
Страница 120: ...UM BIOIIADVANCE DE 1302 17 Abb 1 X Abb 2 X Abb 3 X Abb 4 X...
Страница 139: ...UM BIOIIADVANCE DE 1302 36 11 ZERTIFIKATE 11 1 CE Siehe Anhang A10 1 11 2 GARANTIE Siehe Anhang A10 2...
Страница 140: ...ANNEXOS ANNEXES ANNEXES ANH NGE Bio II Advance...
Страница 141: ...A1 BIOIIADVANCE SP EN 1302 1 Vlam Vent Qext SP Anexo A1 ESQUEMA DE PRINCIPIO EN Annex A1 PRINCIPLE SCHEME...
Страница 142: ...A1 BIOIIADVANCE FR DE 1302 1 Vlam Vent Qext FR Annexe A1 SCH MA DE PRINCIPE DE Anhang A1 FUNKTIONSPRINZIP...
Страница 156: ...A5 1 BIOIIADVANCE FR DE 1302 3 Fig 3...
Страница 183: ...A7 BIOIIADVANCE SP EN FR DE 1302 2...
Страница 184: ...A7 BIOIIADVANCE SP EN FR DE 1302 3...
Страница 185: ...A7 BIOIIADVANCE SP EN FR DE 1302 4...
Страница 186: ...A7 BIOIIADVANCE SP EN FR DE 1302 5...
Страница 187: ...A7 BIOIIADVANCE SP EN FR DE 1302 6...
Страница 188: ...A7 BIOIIADVANCE SP EN FR DE 1302 7...
Страница 189: ...A7 BIOIIADVANCE SP EN FR DE 1302 8...
Страница 190: ...A7 BIOIIADVANCE SP EN FR DE 1302 9...
Страница 191: ...A7 BIOIIADVANCE SP EN FR DE 1302 10...
Страница 226: ......
Страница 227: ......